Guided tutorial
Ideal → mirror → globar → actinic pickup, one source at a time, with your settings restored at the end.
Guided tour. The guided tour runs the simulator in your browser and changes one source at a time, in four steps. Your own settings are restored when it ends.
How the tour works#
- Start. Press Start the guided tour here, or Guided tour in the Explore toolbar. A tour panel opens: beside the plots on a computer, at the bottom of the screen on a phone or tablet.
- Read the step. Each step says what changed and lists numbered things to look at. The same numbers appear on the plots, the source buttons and the view tabs. Tap or click an item to jump to it.
- Continue. Press the highlighted Next button at the bottom of the panel. Back returns to the previous step. On a phone, Hide folds the panel away so you can see more of the plot.
- Leave at any time. End tour (in the panel or the toolbar) or the Esc key ends the tour. Finishing or ending restores your previous controls and selections.
The four steps#
Step 1: Ideal signal
What changed. All five artifact sources are off and the tour defaults are applied. Your own settings are kept and come back when the tour ends.
Where to look
- Spectrum at 100 µs: the artifact-free curve and the effects-applied curve coincide.
- Kinetics at 1697 cm⁻¹: they coincide over the whole decay. The dashed line is the two-exponential + offset projection.
- DAS: the dashed projected spectra follow the input spectra. The next three steps are compared with this.
- Click the contour to choose another time and wavenumber; the other panels follow.
Step 2: Mirror motion
What changed. Mirror motion is on: a 2 nm, 840 Hz sine drive about a 25 nm common bias, temporal phase −100°. All other sources are off.
Where to look
- Contour: broad horizontal bands that change sign with time. The 840 Hz drive (period ≈ 1.19 ms) is sampled by 6 µs bins.
- Kinetics at 1697 cm⁻¹: the effects-applied trace oscillates about the artifact-free decay, by about ±100 µOD on a 600 µOD signal.
- Spectrum at 100 µs: the artifact covers the whole band, including regions with no chemical signal, and grows toward high wavenumber (about 140 µOD at the band edge, against a 460 µOD peak signal).
- DAS: the projected DAS now differ from the input. A fit would report part of the artifact as kinetics.
- Switch Mirror motion off and on to compare.
Step 3: Globar leakage
What changed. Globar leakage is on: continuous leakage at fraction 0.1, blocking OD 0, no changing-leak term. Mirror motion is off again.
Where to look
- Spectrum: the same shape, about 6–12 % smaller depending on wavenumber. A stable leak dilutes the real difference.
- Kinetics: the decay keeps its timing. A stable leak does not create a transient from an unchanged sample.
- Optical contributions, reference intensity: the leak raises the single-beam reference above the ideal IR reference.
- Optical contributions, intensity change: the chemical change ΔS is unchanged; dividing by the larger reference makes ΔA smaller.
Step 4: Actinic (532 nm) pickup
What changed. 532 nm actinic pickup is on: an assumed OPD-modulated return with pulse area 2 × 10⁻⁶ IFG-peak·µs. The selection moves to the 3 µs bin and to the 532 nm alias channel (≈ 1946 cm⁻¹).
Where to look
- Spectrum at 3 µs: a one-channel dip of about −65 µOD at the dotted laser-alias line, where 532 nm light folds onto this sampling grid. Zoom in to see it.
- Kinetics at the alias channel: a spike in the 3 µs bin only; the next bin is clean.
- Contour: the crosshair marks the alias channel at early time. A spectrum taken at 100 µs would miss this response.
Desktop application#
The Windows application runs the same four steps from Guided tour in its header; each press advances one step. Its notes, from TUTORIAL.md:
Click Guided tour in the header. The controls panel is replaced by a tour panel that states what changed at each step and lists numbered places to look; the same numbers mark the plots and source buttons, and clicking an item outlines its plot. Use Next and Back to move between steps. The tour temporarily applies reproducible defaults. Finish or End tour restores previous input controls, time/wavenumber selections, fit-offset choice, selected tab and the exact saved fit (including optimized lifetimes and signed DAS). Theme is retained. Tour steps are single-source examples rather than additions to previously enabled sources. Manual pan/zoom and any edits made during the tour are not saved as a separate project.
- Ideal: inspect 100 µs and 1697 cm−1. The artifact-free and applied curves overlap to roundoff. Click the contour to explore other native bins. The persistent residual is part of the synthetic input, not an instrumental artifact.
- Mirror: inspect a pure 2 nm/840 Hz motion with −100° temporal phase and 25 nm common bias. Compare kinetics and projected DAS. In Demonstrations, compare zero bias and common bias; a pure drive can yield a doubled spectral response.
- Globar: inspect continuous leakage at fraction 0.1 with zero blocking OD and no changing-leak term. Optical contributions shows the reference/intensity change. Stable leakage can dilute true differences; it does not create a transient from an unchanged sample.
- Actinic: inspect the early 3 µs bin at the nominal 532 nm alias near 1945.86 cm−1. An assumed OPD-modulated return is enabled at detector-equivalent pulse area 2 × 10−6 IFG-peak × µs, chosen for readable comparison rather than calibrated laser fluence. Pulse width does not set spatial alias position. A 100 µs slice would miss the prompt response. Direct bypass is a separate physical hypothesis.
- Finish: restore the original workspace. Optional: Demonstrations → Coaddition and phase compares integer-cycle timing, phase walk and 180° pairing. Coadd intensity before taking absorbance; no blanket 1/√N suppression is assumed.
All spectra are synthetic PSI-inspired ΔA. S0 is a separate assumed single-beam reference. No experimental arrays or alternative sample selector are provided. Settings, signs, masks and units are exported; appearance does not change numerical data. The theory reader explains approximation limits and literature evidence. Larger physical displacement phase, unknown response/filtering and actual instrument phase processing require separate validation.