Theory and evidence

From the FringeScope scientific notes. All displayed spectra are simulated; equations and figures are generated from the repository.

New to step-scan artifacts? Take the four-step guided tour in the live simulator, or read the tutorial.

FringeScope is a forward simulator for artifacts in time-resolved step-scan FTIR spectroscopy. It adds declared instrument disturbances to a known synthetic flash-induced difference signal modeled on photosystem I (PSI), so their spectral and kinetic signatures can be compared with the artifact-free input. It is a teaching and experiment-planning tool: all displayed spectra and figures are simulated, and the simulator alone cannot show which mechanism affects a particular measurement. Numerical consistency, literature precedent and instrument calibration are distinct kinds of evidence.

FringeScope
FringeScope magnifying-lens logo
TypeForward simulator of time-resolved step-scan FTIR artifacts
SignalSynthetic photosystem I (PSI)-inspired ΔA: two bin-averaged exponentials plus a causal residual
Spectral grid513 channels, ≈ 2106.39–1053.20 cm⁻¹, descending
Time grid590 bins of 6 µs, −153 to 3381 µs
ArtifactsMirror motion, globar leakage, 532 nm actinic pickup, electrical pickup, return mismatch
Version0.11.0
LicenseMIT

Abbreviations. ΔA: absorbance change (OD). OPD: optical path difference. ZPD: zero path difference. IFG: interferogram. DAS: decay-associated spectrum. HeNe: helium–neon reference laser.

Guided tutorial#

Guided tour. The guided tour runs the simulator in your browser and changes one source at a time, in four steps. Your own settings are restored when it ends.

Start the guided tour

Use Guided tour in the header to follow ideal → mirror → globar → actinic leakage. A tour panel states what changed and numbers the plots to inspect; Next and Back move between steps, and Finish or End tour restores the previous controls and selections. The tour temporarily uses declared default parameters. It does not change the synthetic dataset or theme. TUTORIAL.md contains the same sequence and an optional electrical/coaddition lesson.

  • Ideal: inspect the 100 µs spectrum and 1697 cm−1 kinetics. Reference and effects-applied curves overlap. Move to a late time to see the unresolved residual; its presence is part of the input model, not an artifact.
  • Mirror: inspect the broad oscillation produced by a 2 nm, 840 Hz drive about a 25 nm common bias. Compare the input and projected DAS. Turn the source off/on to compare. Open Demonstrations → Mirror harmonics and PSI to distinguish mechanical frequency from spectral harmonics.
  • Globar: the tour isolates continuous leakage. Compare Optical contributions with the ideal reference. A stable leak can dilute a real difference without creating a new transient from an unchanged sample.
  • Actinic: the tour isolates an assumed OPD-modulated 532 nm return and selects the early 3 µs bin near its predicted alias. Its pulse is brief; the 100 µs spectrum would miss this prompt response. A direct bypass is a separate path with no wavelength-specific carrier.
  • Finish: restore the original workspace. Optional: open Coaddition and phase to see why timing, rather than record count alone, controls coherent artifact suppression.

The simulated PSI signal#

Artifact-free PSI ΔA is a signed transient difference signal. Explore and Demonstrations use the same generator. S0 is a separate positive single-beam intensity envelope used for reconstruction and ratio formation; it is not sample absorbance or a calibrated detector/filter response.

Delta A(nu,t) = [H(nu) exp(-t/tau_H) + R(nu) exp(-t/tau_R) + C(nu)] u(t)
(1)

H and R are heating-like and recombination-like amplitude spectra. C is a causal unresolved residual; u(t) is zero before excitation and unity afterwards. The model uses ordinary exponentials with displayed lifetimes τH = 12.16 µs and τR = 449.14 µs, with βH = βR = 1. These are illustrative source parameters, not newly measured lifetimes or uncertainty estimates. Calculations and exports retain their precision.

Recorded bin mean equals the time integral over the bin divided by Delta t.
(2)

Chemical bases are averaged over 6 µs recorder bins. Centers run from −153 to 3381 µs. The 100 µs spectrum is linearly interpolated between centers at 99 and 105 µs. Selection clicks snap to native channels/bins. Wavenumber remains descending, from about 2106.39 to 1053.20 cm−1 across 513 channels. The acquisition origin and periodic 1024-point interferogram are model conventions, not a reconstruction of an OPUS experiment.

The recombination approximation has declared archival influence but contains no measured spectral matrix. H uses six Gaussian centers guided by Nelson et al. (2026), §6.7/Figs. 7–9: 1662 positive, 1623 negative, 1561 negative, 1545 positive and 1473 negative cm−1, plus a weak negative feature near 1699 cm−1. Widths and amplitudes are chosen sensitivity settings; no published curve is digitized or fitted. The cited paper qualifies the heating assignment and does not calibrate this model's temperature rise.

C is defined as C(ν̃) = 0.08 R(ν̃) + 0.04 H(ν̃). Those fractions are assumed, not measured yields or a determined slow state. H, R and C scale independently. Fitting a constant changes the analysis basis, not the input residual. Component functions retain a declared linear edge taper: zero outside 1120–1950 cm−1 with 35 cm−1 transitions. This is a chemical model boundary, not measured filter transmission. No smoothing, baseline correction or per-spectrum normalization is applied.

Frequency, phase and spectral shape#

At 840 Hz the period is 1.19 ms, giving about 2.84 cycles in 3.38 ms. Temporal phase shifts maxima without changing the period. An even response to displacement can produce 2f from a pure sine drive; common bias can restore a strong f term. Composite drive harmonics and nonlinear readout harmonics are different mechanisms.

Period is 1/f; cycle count is f times the time window; phase timing shift is Delta phi/(360 degrees times f).
(3)
For a small signal, Delta A approximately equals -Delta S divided by ln(10) S0.
(4)

A broad additive intensity error may be amplified in absorbance where S0 is small. That does not uniquely identify its cause. Uniform coherent mirror displacement in the ideal biased real-readout model instead has an approximately wavenumber-squared dependence at small phase. Adding temporal harmonics alone does not guarantee an inverse-transmission spectral shape. A short oscillatory record cannot uniquely identify mechanical modes or a mains source; 14 × 60 = 840 Hz is only an arithmetic coincidence.

Mirror displacement and detector response#

Interferogram I(x,t) is the cosine transform of single-beam spectrum S(nu,t).
(5)
OPD displacement is twice mirror displacement, in matching length units. Shift is expanded through fourth order.
(6)

OPD x is in cm; mirror displacement d in nm changes OPD by 2d × 10−7 cm. The production engine uses a fourth-order expansion at the physical optical wavenumber, retaining the alias sign of the [B, 2B] display zone. Common bias affects both records; return mismatch and per-step jitter are differential errors. Constant ZPD registration error alone is not a temporal 840 Hz source. A step-scan mirror normally holds a position during a flash train, rather than returning to ZPD after every flash.

A small mirror displacement d changes the AC detector signal by twice d times the local OPD derivative about common mirror bias b.
(7)

The reference interferogram's local slope controls the first-order detector change. At an unbiased symmetric ZPD that derivative vanishes, so higher orders matter. The lesson below evaluates an exact spatial cosine sum and displays the AC detector component, not unmodulated detector DC. Baseline time zero is an arbitrary acquisition origin, not a flash. All traces share one ideal IR IFG peak; stated vertical offsets are graphical only.

Detector changes at native OPDs
Figure 1. Original simulated OPD detector traces and their reference/slope. Solid traces use exact uniform displacement; dashed traces use the first-order local slope. There is no excitation or random noise. Parameters, units and offsets appear in the figure. Inspired by detector-position examples in Rödig and Siebert (1999), not a reproduction of their data.

Harmonics and coaddition#

For fixed real readout and stationary biased reference, difference absorbance is minus log10 of cos(kappa times bias plus drive) divided by cos(kappa bias).
(8)

Here a and b are mirror amplitude and common bias in nm, θ is temporal drive phase and κ = 4πν̃ × 10−7 cm nm−1 with ν̃ in cm−1. The exact uniform fixed-real identity explains leading 2f at zero bias and a strong fundamental with bias. The stationary reference shares that bias. This is not the instrument's phase-correction algorithm.

Mirror harmonics and the shared PSI signal
Figure 2. Pure sine motion with zero/common bias, harmonic amplitudes over one exact cycle, and superposition on the shared PSI input. DC is excluded from harmonic bars; amplitudes below the indicated floor remain in numerical exports. The chemical panels retain bin-averaged chemical bases and flash-gated, bin-center motion. Their timing differs explicitly from the ungated baseline lesson.
Average the N detector or intensity records first, then form difference absorbance against the declared reference.
(9)
Intensity harmonic m has phase-coherence factor C_m; a 0 and pi phase pair cancels odd orders and retains even orders.
(10)

Detector/intensity records are averaged before the logarithm. An ideal 0°/180° pair cancels odd intensity harmonics while retaining even ones. At 840 Hz and 10 Hz repetition, an exact integer 84 cycles between shots preserves coherence; 10.01 Hz gives phase walk. Cm describes intensity-harmonic transfer, not universal absorbance attenuation. Coherent OPD phase is imposed in this lesson to isolate shot timing; no free-running step-phase schedule or 1/√N suppression is assumed.

Timing-dependent coherent averaging
Figure 3. Original timing/coaddition lesson with eight-record traces and harmonic bars. The source paper discusses doubled-frequency survival under phase-correlated averaging on pp. 897–898. The plotted case is nonlinear mirror response: a purely linear additive electrical sine would cancel under ideal pairing and needs an even-harmonic/nonlinear contribution to show survival.

Optical folding and pulsed pickup#

Spatial sampling frequency Fs is 1/Delta x = 2B; desired band is B to 2B.
(11)
q(nu) = abs(((nu + B) modulo 2B) - B); displayed wavenumber is 2B - q.
(12)

Spatial sampling, not nanosecond pulse width, sets the optical alias. For this grid, displayed in the zone [B, 2B], B ≈ 1053.20 cm−1 and OPD step ≈ 4.75 µm, nominal 532 nm OPD-modulated light maps to about 1945.86 cm−1. Directly sampled fringe checks support the identity; these values are not measured metrology or detector-response calibration. A uniform direct bypass has no optical carrier; an assumed step-dependent bypass has the spectrum of its coupling pattern.

The optical pulse model is a nominal 6 ns Gaussian plus an assumed 20 ns delay tail, convolved with a first-order detector response and integrated over recorder bins. A 20 ns free-space delay corresponds to about 6 m additional path; short reflections do not automatically justify it. A Gaussian centered at zero places some area before zero, including the bin centered at −3 µs. Electronic recovery is a separate assumed response. Pulse-area settings are detector-equivalent quantities, not calibrated laser fluence.

Continuous leakage and electrical pickup#

Observed Delta A = -log10[(S0 times 10 to minus true Delta A + leakage L)/(S0 + L)].
(13)
For zero true change, the same steady leakage in both records gives zero observed difference absorbance.
(14)
For a small difference signal, the observed amplitude is scaled by S0 divided by S0 plus steady leakage.
(15)

Stable positive leakage L enters reference and signal, diluting a small real difference while canceling for an unchanged sample. Globar leakage uses Planck radiance per wavenumber, an assumed Gaussian transmission leak and flat responsivity. Changing leakage, saturation and calibrated shot noise are not inferred from this steady model. HeNe is not a primary control; its internal comparison demonstrates stationary cancellation and a conditional mirror-coupled exception, not harmlessness in the real instrument.

Conditional CW contribution
Figure 4. Original steady-CW comparison and conditional motion counterexample, on separately labeled scales. The nominal wavelength, assumed coupling and fixed bias/motion are recorded in the source metrics. No spectra are normalized or smoothed. Visible response and actual optical path remain unmeasured.

Electrical pickup is an additive detector term with an explicit spatial coupling. The example below assumes Gaussian coupling near ZPD; a uniform coupling would transform differently. The IR reference remains positive. A blocked/zero reference cannot define an absorbance ratio, even though IR-blocked detector controls are useful experimentally.

Mirror and electrical baseline comparison
Figure 5. Original unexcited mirror/electrical maps and native-time slices. Each map has its own labeled signed scale. Acquisition zero is an arbitrary oscillator origin. The paper's imposed ±4 nm at 50 Hz differs from these illustrative 2 nm/840 Hz settings. Similar-looking oscillations do not establish identical mechanisms.

Numerical domain and evidentiary limits#

The displacement audit compares the fourth-order engine with an independent physical-wavenumber cosine sum at OPD-dependent shifts. In the tested spectrum/pattern, maximum spectral discrepancy is about 0.03 µOD at 0.10 rad, 2.86 µOD at 0.25 rad and 87.32 µOD at 0.50 rad. The 0.25 rad warning is a caution threshold, not an accuracy guarantee or a uniform bound for all settings. Low transmission and higher physical optical frequencies require separate checks.

Spatial expansion convergence
Figure 6. Original convergence audit for a specified OPD-dependent displacement pattern. The statistic is maximum interferogram discrepancy divided by ideal IFG peak; it is not a measurement uncertainty. Full settings, source hashes and numerical arrays are recorded by the audit script.

Motion/electrical terms are evaluated at recorder-bin centers. A pure 840 Hz sine loses about 0.004% amplitude under a 6 µs box average; the difference grows at higher frequencies and does not bound nonlinear motion/chemical products. Independent quadrature checks the pulse PDF against CDF bins for selected cases; agreement validates implementation consistency, not the assumed physical response.

The full mathematical FFT band is approximately 2106.39–1053.20 cm−1. Makita's dissertation reports an acquired range approximately 2106.39–1065.54 cm−1; these are not identical grids. The periodic phase-zero reconstruction does not reproduce an original single-sided acquisition, instrument line shape or OPUS phase algorithm. Nominal visible wavelengths require consistent air/vacuum and metrology conventions.

Literature and assumptions#

  • Rödig and Siebert (1999), pp. 895–899: primary evidence for mirror/electrical artifacts and averaging dependence. Their experiments do not identify this simulator's 840 Hz source. DOI.
  • Makita (2018), printed p. 29/PDF p. 76: describes folding precautions, two 2000–1000 cm−1 filters and CaF2 windows. These are reported apparatus precautions, not measured zero-leakage transfer curves. Institutional record.
  • Makita and Hastings (2016), pp. 1467–1468: describes filter placement. DOI.
  • Nelson et al. (2026), §6.6–6.8/Figs. 7–9: supports conditional PSI heating-like features and the need to qualify unresolved slow components. The present two-plain-exponential-plus-offset model is different from that published fit. DOI.
  • Manning and Griffiths (1997): cited for low-frequency fluctuations and noise sources in step-scan measurements; no quantitative parameters are taken from it. DOI.

Absolute amplitudes, thermal fractions, actual detector/filter response and the disturbance mechanism remain assumptions or unmeasured quantities. Synchronized position-error records, raw interferograms, IR/laser-blocked controls and timing/coaddition tests are needed for instrument-specific discrimination. Numerical fits, spectral similarity and inverse-transmission enhancement alone do not identify a cause.

Fitting, exports and reading#

Ordinary updates project an unweighted fixed kinetic basis. Refit optimizes log lifetimes with signed amplitudes and β = 1; the default optimization region is 0–3378 µs and 1200–1900 cm−1. Bound hits and unresolved lifetimes are flagged. Optimizer success does not establish a physical assignment, SEM or confidence interval. Artifact-biased DAS are projected components, not necessarily molecular spectra.

References below 1% maximum and nonpositive reconstructed intensities are masked, never floored into absorbance. Full grids/masks remain in NPZ; DPT retains channels finite at every time and records exclusions/units. New export folders contain exact settings and synthetic parameters. Publication figures stay light; screen captures record the selected appearance. Dark/light selection does not change numerical data.

The offline reader defaults to dark, uses the available width, and retains proper indices/powers. Image paragraphs use single spacing so captions follow figures; text size is adjustable. Demonstrations use fixed acquisitions independent of workspace toggles. Their source scripts reproduce the released synthetic figures without measured arrays.

Model assumptions and provenance#

The runtime source is one analytic ΔA matrix inspired by photosystem I (PSI) difference spectra. The main workspace and demonstration chemical panels share its generator. Positive/negative amplitudes are retained. The reference S0 is a distinct assumed single-beam envelope with a positive default minimum about 0.41 in arbitrary intensity units, not measured sample absorbance or detector response.

The source lifetimes are displayed as 12.16 and 449.14 µs; full values and coefficients are in synthetic_model.json and exports. β = 1. A causal residual is C = 0.08 R + 0.04 H. The fractions, Gaussian widths/amplitudes and persistence are sensitivity assumptions. Heating centers use the signed PSI features described by Nelson et al. (2026), §6.7/Figs. 7–9. The recombination parameterization has declared archival influence; no measured arrays are included. The current residual is not a determined third molecular state or calibrated thermal tail.

Time centers span −153–3381 µs with 6 µs spacing. Chemical bases are bin averaged; 100 µs spectra use labeled linear interpolation. Motion/electrical terms are evaluated at bin centers. Source functions have a linear 35 cm−1 edge taper and are zero outside 1120–1950 cm−1. No runtime smoothing, baseline correction or amplitude normalization is applied. Paired axis reversal is used internally only when required for interpolation; stored/displayed/exported wavenumbers remain descending.

The mathematical 513-channel band spans about 2106.39–1053.20 cm−1. The 1024-point periodic, phase-zero interferogram is a model construction, not an OPUS raw acquisition. The dissertation's reported acquired endpoint near 1065.54 cm−1 is distinct. The reference threshold is 1% maximum. Nonpositive reconstructed intensities remain masked; no positive floor is substituted.

The production displacement engine uses a fourth-order physical-wavenumber expansion. Its 0.25 rad warning is not a universal accuracy guarantee: the bounded audit found spectral discrepancies about 2.86 µOD at 0.25 rad and 87.32 µOD at 0.50 rad for one OPD-dependent pattern. Exact uniform-displacement lessons are separate from the OPD-dependent approximation and actual instrument phase correction. Magnitude and fixed-real readouts are sensitivity models.

Continuous optical leakage is included in reference and signal. It can dilute a real difference but cancels for a stationary unchanged sample. Visible responsivity, path/coherence, filter response and source power remain unmeasured. An optical-wavelength alias requires OPD-dependent coupling; uniform direct bypass does not have that carrier. Pulses use an assumed Gaussian/tail/electronics response integrated through bins. A pulse leading half can enter the −3 µs bin. Temporal response validation checks numerical consistency, not physical calibration.

The electrical example's spatial coupling is assumed. 14 × 60 = 840 Hz does not identify a source. Coaddition is performed before the logarithm; harmonic coherence depends on shot/step timing. No blanket 1/√N artifact suppression is assumed. Fixed demonstrations impose coherent OPD phase and state their references, AC detector normalization and graphical offsets.

Unweighted fits project signed amplitudes; free fits optimize log lifetimes with β fixed. No SEM/confidence interval is asserted. Numerical recovery and spectral similarity do not establish physical assignments. Full settings/grids/masks remain in exports. Dark/light appearance does not alter data. Original application code and synthetic inputs use the MIT license. Clean-second-host verification and complete binary corresponding-source/build-input verification are not established by numerical tests.

References#

  1. Thorlabs. Laser and ASE Systems: red helium-neon lasers. https://www.thorlabs.com/catalogpages/571.pdf Nominal 632.8 nm specification only; does not establish the user's wavelength, detector response or leakage
  2. Rödig, C.; Siebert, F. (1999). Errors and Artifacts in Time-Resolved Step-Scan FT-IR Spectroscopy. Applied Spectroscopy 53, 893–901. doi:10.1366/0003702991947595
  3. Manning, Christopher J.; Griffiths, Peter R. (1997). Noise Sources in Step-Scan FT-IR Spectrometry. Applied Spectroscopy 51, 1092–1101. doi:10.1366/0003702971941755
  4. Makita, Hiroki (2018). Bioenergetics in Photosystem I: Time-Resolved Step-Scan FTIR and Visible Spectroscopic Studies of the Secondary Electron Acceptor A1. Georgia State University. https://scholarworks.gsu.edu/phy_astr_diss/106 Folding and filter description: printed p. 29, PDF p. 76
  5. Makita, Hiroki; Hastings, Gary (2016). Time-resolved visible and infrared absorption spectroscopy data obtained using photosystem I particles with non-native quinones incorporated into the A1 binding site. Data in Brief 7, 1463–1468. doi:10.1016/j.dib.2016.04.031
  6. Nelson, Michael Raymond; Tamanna, Sabiha; Asjad, Muhammad; Hastings, Gary (2026). A Reproducible Two-Stage Protocol for Global Fitting of Time-Resolved Fourier Transform Infrared Difference Spectra. Methods and Protocols 9, 158. doi:10.3390/mps9060158